Contacts
-
Instrument Scientist
-
Instrument Scientist
-
Instrument Scientist
-
Neutron Scattering Scientist/Group Leader
Neutron reflectometry probes surface and interfacial structures of thin films on length scales of 0.5 nm to 350 nm.
The Liquids Reflectometer measures specular and off specular neutron reflectivity in a horizontal sample geometry from solid surfaces, tilted solid/liquid interfaces, and free liquid surfaces. Probe wavelength and flux constrain size and flatness to samples no smaller than about 1×1 cm2 deposited on atomically smooth and cm-scale flat substrates such as silicon, quartz, sapphire, or water. The most commonly used substrate is a 50-mm-diameter, 5-mm-thick silicon wafer. A variety of sample environments are supported, including an 18-slot robotic sample changer, temperature-controlled liquid/solid and potentiostatically controlled electrochemical cells, vacuum and gas-handling chambers, a rheometer, and a Langmuir trough. For specialized sample environments, the user should contact a beamline staff member.
Researchers are using the LR to study a broad range of problems, including those in the following scientific fields
Moderator | Coupled supercritical H2 |
Bandwidth | 3.4 Å @ 60 Hz, 6.8 Å @ 30 Hz |
Wavelength Range | 2.5 Å < λ < 17.0 Å |
Resolution | 0.02 < δQ/Q < 0.10 |
Sample Environment | 18 slot sample-changing robot 2”, 3”, and 4” diameter liquid/solid cells Potentiostat Langmuir trough Vacuum and multi-environment chambers Rheometer |
Instrument Scientist
Instrument Scientist
Instrument Scientist
Neutron Scattering Scientist/Group Leader