Enclosed Balance

Description: Enclosed Balance Minimum capacity: 0.0001 g Max capacity: 21 g
Resource Type: Balances
Location ID: 8700, TA-B106-B
Facility: SNS
Lab Space Manager: Alex Koldys
Specifications: Verified models with EC type-approval Model LE26P-OCE LE225D-OCE LE324S-OCE LE244S-OCE Type BC BL 100 BC BL 100 BD BL 100 BD BL 100 Accuracy class1) k k k k Weighing capacity, Max1) g 5/21 100/220 320 240 Scale interval d1) mg 0.002/0.01 0.01/0.1 0.1 0.1 Tare range (subtractive) g < 100% of the maximum capacity Verification scale interval e1) g 0.0001 0.001 0.001 0.001 Minimum capacity, Min1) g 0.0002 0.001 0.01 0.01 Response time (average) s 10 < 6/3 < 3 < 2 Range of use according to CD1) g 0.0002–21 0.001–220 0.01–320 0.01–240 Allowable ambient operating temperature: – with »isoCAL« function °C 0° to +40° (+32° to +104°F) – without »isoCAL« function °C +15° to +25° (+59° to +77°F) External calibration weight g 20 (E2) 200 (E2) 200 +100 (E2) 200 (E2) (of at least accuracy class…) Net weight, approx. kg 7.6 7.6 6.5 6.5 Weighing pan size mm 50 d 80 d* 80 d* 80 d* (inner diameter) Weighing pan surface cm2 20 64* 64* 64* Weighing chamber height (weighing pan to draft shield cover) mm 162 232 232 232 Dimensions (W+D+H) – Balance mm 213+342+270 213+342+340 213+342+340 213+342+340 – Electronics box mm 134+51+155 134+51+155 – – AC power source/ V~ AC adapter STNG6, 230 V or 115 V, power requirements +15%…– 20% (protection rating IP20) Frequency Hz 48 – 60 Power consumption (average) VA maximum 16; typical 8 Approx. hours of operation with the YRB05Z reachargeable battery pack h 20 20 22 22 Selectable weight units Grams, carats, milligrams Built-in interface RS-232C-S/V24-V28; 7-bit; parity: even, odd, mark, or space; transmission rates: 150 to19,200 baud; 1 or 2 stop bits; software/hardware handshake
Details: Manufacturer:SARTORIUS , Model:LE225D
Attachments: Manual, Specifications
Access Requirements:

Scientific Laboratory Access Training for Neutron Sciences Users, RSS Required Reading, Rad Worker I or II, and Site-Specific Training
All work, storage, and removal of samples requires lab space manager and radiological control technician (RCT) approval


Alternate Contacts: Luke Heroux, Maya C Little, Qiu Zhang, Kristin Nevius